Coolline Model CLS301/400/600 Instruction Manual - Third Edition - Yamato Scientific Co. LTD. z Thank you for purchasing "Coolline, CLS Seri
7 Before Using This Unit Installation Procedure 1 Release the stopper lock of the casters. (CLS400/600) Push down the stopper button of the cast
8 Before Using This Unit Installation Procedure 6 Connecting the power. Confirm that the leakage breaker and the POWER switch are turned "O
9 Description and Function of Each Part Main Unit Rear view Front view Control panel Drain cock Earth leakage breaker Caster (Two front ca
10 Description and Function of Each Part Control Panel No. Name Function ① RUN/STOP Key Used for operation Start/Stop. ② ▲▼ Key Selects sett
11 Description and Function of Each Part Operation Monitor No. Name Function ① REFRIGERATOR ERROR Lamp Lights when the refrigerator is
12 Description and Function of Each Part Characters of the Controller The characters controller shows are as follows: Character Identifier Name
13 Operation Method Operation Mode and Function List The operation modes of this unit are as follows; Name Description PageFixed Temperature Opera
14 Operation Method Operation Mode and Function List The operation functions of this unit are as follows; Name Description PageCalibration offset
15 Operation Method Operation Mode, Function Setting Key, and Characters The operation mode setting and function setting use the key operation and
16 Operation Method Fixed Temperature Operation In this mode, the unit starts to operate by pressing RUN/STOP key and continues operating at the se
Contents Cautions in Using with Safety...1 • Explanation...
17 Operation Method Fixed Temperature Operation 3. Start operation • Press RUN/STOP key for a second. • FIXED TEMP. lamp will be lit and operat
18 Operation Method Quick Auto Stop Operation Quick auto stop operation procedure This operation is used to specify the period up to automatic
19 Operation Method Quick Auto Stop Operation 4. Stop/terminate timer operation • The operation will be stopped automatically at the setting tim
20 Operation Method Auto Stop Operation In this mode, the unit automatically comes to a stop after the set period passes away from the start of fix
21 Operation Method Auto Stop Operation 3. Select auto stop operation • Press TIMER key, and display the character "AStP" which means
22 Operation Method Auto Stop Operation 6. Stop/terminate timer operation • The operation will be stopped automatically at the setting time. •
23 Operation Method Auto Start Operation In this mode, the unit automatically starts to operate after the set period passes away from the start of
24 Operation Method Auto Start Operation 3. Select auto start operation • Press TIMER key, and display the character "AStr" which mean
25 Operation Method Auto Start Operation 6. Stop/terminate timer operation • After timer operation, it will start operation at the setting time.
26 Operation Method Calibration Offset Function Calibration offset is a function which corrects the difference between the temperature in bath and
Specification...45 Wiring Diagram...
27 Operation Method Lock Function Lock function that makes operation setting unchangeable. ① Press SUB MENU key for 4 seconds. Then by pressing
28 Operation Method Power Failure Compensation Function This is the setting that can start the operation with the former setting in case of electri
29 Operation Method Addition Time Function Displays the time length that the controller is plugged. ① Press SUB MENU key for 4 seconds. Then by
30 Operation Method Temperature Output Terminal Precautions • Operate this product according to the procedure described in this instruction manu
31 Operation Method Temperature Output Terminal Specification Temperature Output (ANALOG) • The voltage (DC) corresponding to the measured temper
32 Operation Method Cooling curve, cooling capacity curve (reference data) The graphs show the cooling and cooling capacity curves of each model
33 Operation Method Cooling curve, cooling capacity curve (reference data) CLS400冷却曲線 室温20℃-20.0-10.00.010.020.030.00 30 60 90 120経過時間(分)温度(℃)外部槽液
34 Operation Method Cooling curve, cooling capacity curve (reference data) CLS600冷却曲線 室温20℃-20.0-10.00.010.020.030.00 30 60 90 120経過時間(分)温度(℃)外部槽液
35 Operation Method Cooling curve, cooling capacity curve (reference data) CLS冷却能力曲線01002003004005006007008009001000-40 -30 -20 -10 0 10 20 30 40
36 Operation Method Nybrine freezing temperature and viscosity (reference data) 0-10-20-30-40-500 1020304050607080901000-10-20-30-40-50ナイブライン濃度(wt
1 Cautions in Using with Safety Explanation MEANING OF ILLUSTRATED SYMBOLSVarious symbols are used in this safety manual in order to use the unit
37 Handling Precautions WARNING! If a problem occurs If smoke or strange odor should come out of this unit for some reason, turn off the POWER sw
38 Handling Precautions Countermeasure for stop operation during night or long-term stop In case of stopping operation during night or long-term,
39 Maintenance Method Daily Inspection and Maintenance For the safety use of this unit, please perform the daily inspection and maintenance without
40 Maintenance Method Daily Inspection and Maintenance Cleaning the filter The mesh plate is fixed with a magnet. Pull it toward you. The bottom of
41 Long storage and disposal When not using this unit for long term / When disposing CAUTION! When not using this unit for long term… • Turn off t
42 In the Event of Failure… Safety Device and Error Code This unit has an automatic diagnosis function built in the controller and safety devices i
43 In the Event of Failure… Trouble Shooting Phenomenon Check point The unit does not start to operate although the earth leakage breaker and POW
44 After Service and Warranty In Case of Request for Repair If the failure occurs, stop the operation, turn OFF the POWER switch, and unplug the
45 Specification Product Name Coolline Model CLS301 CLS400 CLS600 Circulation unit Circulation in the external open system Usable ambient temp
46 Wiring Diagram CLS301 1T1ELB2354AC100V6X1X2PLB12TB1PIO120CN1CN120134567891011TB2121314151617X3-2X2PSWX3-1PSX3X4-1FSX4X4-221+-FM21433123RSCOVRX5
2 Cautions in Using with Safety Table of Illustrated Symbols Warning Warning, generally Warning, high voltage Warning, high temperatureWarnin
47 Wiring Diagram CLS400 1T1ELB2354AC100V6X1X2PLB12TB1PIO120CN1CN120134567891011TB2121314151617X3-2X2PSWX3-1PSX3X4-1FSX4X4-2+-FM21433123RSCOVRX5C1
48 Wiring Diagram CLS600 FMRSCOVRRF1T1ELB2354AC100V6X1X2PLB12TB1PIO120CN1CN120134567891011TB2121314151617X3-2X2PSWX3-1PSX3X4-1FSX4X4-2+-2143X5C1C2
49 Replacement Parts Table Common Parts Symbol Part Name Code No. Specification Manufacturer WIB Operation display board LT00006042 Toho D
50 Reference List of Dangerous Substances Never use explosive substances, flammable substances and substances that include explosive or flammable
51 Installation Standard Manual Install the unit according the procedure described below (check options and special specifications separately). M
52 Responsibility Please follow the instructions in this document when using this unit. Yamato Scientific has no responsibility for the accidents
3 Cautions in Using with Safety Fundamental Matters of "WARNING!" and "CAUTION!" WARNING! Do not use this unit in an area wh
4 Before Using This Unit Requirements for Installation 1. Always ground this unit • Be sure to connect the earth wire (the green cable of power
5 Before Using This Unit Requirements for Installation 4. Do not modify • Modification of this unit is strictly prohibited. This could cause a fa
6 Before Using This Unit Requirements for Installation 8. Handling of power code • Do not entangle the power cord. This will cause overheating an