Coolline Model CLH301/400/600 Instruction Manual - Second Edition - Yamato Scientific Co. LTD. z Thank you for purchasing "Coolline, CLH Ser
7 Before Using This Unit Installation Procedure 1 Release the stopper lock of the casters. (CLH400/600) Push down the stopper button of the cast
8 Before Using This Unit Installation Procedure 6 Connecting the power. Confirm that the leakage breaker and the power switch are turned "o
9 Description and Function of Each Part Main Unit Rear view Front view Control panel Drain cock Earth leakage breaker Caster (Two front ca
10 Description and Function of Each Part Control Panel No. Name Function ① RUN/STOP Key: Starts/stops the operation. ② ▲▼ Key: Uses for rising
11 Description and Function of Each Part Operation Monitor No. Name Function ① Refrigerator Error Lamp Lights when the refrigerator is
12 Description and Function of Each Part Characters of the Controller The characters controller shows are as follows: Character Identifier Name
13 Operation Method Operation Mode and Function List The operation modes of this unit are as follows; Name Description PageFixed Temperature Oper
14 Operation Method Operation Mode and Function List The operation functions of this unit are as follows; Name Description PageAuto overheating p
15 Operation Method Operation Mode, Function Setting Key, and Characters The operation mode setting and function setting use the key operation and
16 Operation Method Setting of Overheating Prevention Device The unit has the overheating prevention device (manual reset) that consists of indepen
Contents Cautions in Using with Safety...1 • Explanation...
17 Operation Method Fixed Temperature Operation In this mode, the unit starts to operate by pressing RUN/STOP key and continues operating at the se
18 Operation Method Fixed Temperature Operation 4. Start operation • Press the orange RUN/STOP key for about one second. The unit starts opera
19 Operation Method Quick Auto Stop Operation Quick auto stop operation procedure This operation is used to specify the period up to automatic
20 Operation Method Quick Auto Stop Operation To correct or check setting… Changing the setting temperature during operation is possible by pressi
21 Operation Method Auto Stop Operation In this mode, the unit automatically comes to a stop after the set period passes away from the start of fix
22 Operation Method Auto Stop Operation 2. Start timer operation • Press the RUN/STOP for one second after deciding the time. • Timer operatio
23 Operation Method Auto Start Operation In this mode, the unit automatically starts to operate after the set period passes away from the start of
24 Operation Method Auto Start Operation 2. Start timer operation • Press the RUN/STOP for one second after deciding the time. • Timer operati
25 Operation Method Calibration Offset Function Calibration offset is a function which corrects the difference between the temperature in bath and
26 Operation Method Lock Function This function locks the operation status previously set. The function can be set or cancelled by the SUB MENU
Wiring Diagram...55 • CLH300 ...
27 Operation Method Temperature Output Terminal Precautions • Operate this product according to the procedure described in this Operation Manual
28 Operation Method Temperature Output Terminal Specification Temperature Output (ANALOG) • The voltage (DC) corresponding to the measured temper
29 Operation Method RS485 Communication Function 1. Settings Relating to Communication 1.1 Communication Settings Before starting communication wi
30 Operation Method RS485 Communication Function 2. Data Transmission Method Item Specification Communication standard EIA standard, complying wi
31 Operation Method RS485 Communication Function 4.2 Message Types Message types include transmission request messages from the host computer an
32 Operation Method RS485 Communication Function 4.4 Reply Message Structures 4.4.1 Reply Messages to Read Request Messages S T X □ □ACK □ □ □
33 Operation Method RS485 Communication Function 4.5 Description of Codes The following codes from ①STX, ②address to ⑩error type are represen
34 Operation Method RS485 Communication Function ⑨ NAK This is a negative acknowledgement code and included and returned in the "reply messag
35 Operation Method RS485 Communication Function 5. Communication Examples 5.1 Read communication example Example) Request message: A request for
36 Operation Method RS485 Communication Function 5.2 Write communication example Example) Request message: A request for setting "SV to 135&
1 Cautions in Using with Safety Explanation MEANING OF ILLUSTRATED SYMBOLSVarious symbols are used in this safety manual in order to use the unit
37 Operation Method RS485 Communication Function 6. Wire Connection Shown below is an example of multi-drop wire connection. PC VS3 controller CN
38 Operation Method RS485 Communication Function 7. List of Identifiers/Commands <Identifiers and set values> *1: "_" means a spac
39 Operation Method RS485 Communication Function Other Parameters Name Identifier Command Setting Value Key lock LOC R/W 00000:Key lock release
40 Operation Method RS485 Communication Function 8. List of ASCII Codes ASCII code 02H 03H 06H 15H Symbol STX ETX ACK NAK ASCII code
41 Operation Method Cooling curve, cooling capacity curve (reference data) The graphs show the cooling and cooling capacity curves of each model
42 Operation Method Cooling curve, cooling capacity curve (reference data) CLH400加熱・冷却曲線 室温20℃-20.0-10.00.010.020.030.040.050.060.070.080.090.0100
43 Operation Method Cooling curve, cooling capacity curve (reference data) CLH600加熱・冷却曲線 室温20℃-20.0-10.00.010.020.030.040.050.060.070.080.090.0100
44 Operation Method Cooling curve, cooling capacity curve (reference data) CLS冷却能力曲線01002003004005006007008009001000-40 -30 -20 -10 0 10 20 30 40
45 Operation Method Nybrine freezing temperature and viscosity (reference data) 0-10-20-30-40-500 1020304050607080901000-10-20-30-40-50ナイブライン濃度(wt
46 Handling Precautions WARNING! If a problem occurs If smoke or strange odor should come out of this unit for some reason, turn off the power ke
2 Cautions in Using with Safety Table of Illustrated Symbols Warning Warning, generally Warning, high voltage Warning, high temperatureWarnin
47 Handling Precautions Recovery from a power failure If the unit was deactivated in the middle of operation due to a power failure and is re-ene
48 Maintenance Method Daily Inspection and Maintenance For the safety use of this unit, please perform the daily inspection and maintenance without
49 Maintenance Method Daily Inspection and Maintenance Cleaning the filter The mesh plate is fixed with a magnet. Pull it toward you. The bottom of
50 Long storage and disposal When not using this unit for long term / When disposing CAUTION! When not using this unit for long term… • Turn off t
51 In the Event of Failure… Safety Device and Error Code This unit has an automatic diagnosis function built in the controller and safety devices i
52 In the Event of Failure… Trouble Shooting Phenomenon Check point The unit does not start to operate although the leakage breaker is turned on.
53 After Service and Warranty In Case of Request for Repair If the failure occurs, stop the operation, turn OFF the power switch, and unplug the
54 Specification Product Name Coolline Model CLH301 CLH400 CLH600 Circulation unit Circulation in the external open system Usable ambient temp.
55 Wiring Diagram CLH301 AC100V1SSR2134T1ELBX2X22345786CT1①②③④⑤⑥⑦⑦⑧⑩⑱⑯⑰PLBWIB1・・・・・・・・・23CN512CN4CN223121CN121CN74365X1KPIOTB121118CN8CN11183TB221
56 Wiring Diagram CLH400 AC100V1SSR2134T1ELBX2X22345786CT1①②③④⑤⑥⑦⑦⑧⑩⑱⑯⑰PLBWIB1・・・・・・・・・23CN512CN4CN223121CN121CN74365X1KPIOTB121118CN8CN11183TB221
3 Cautions in Using with Safety Fundamental Matters of "WARNING!" and "CAUTION!" WARNING! Do not use this unit in an area wh
57 Wiring Diagram CLH600 ELBAC100V⑪⑫⑬⑭FMRSCOVRRF1SSR2134T1X2X22345786CT1①②③④⑤⑥⑦⑦⑧⑩⑱⑯⑰PLBWIB1・・・・・・・・・23CN512CN4CN223121CN121CN74365X1KPIOTB121118C
58 Replacement Parts Table Common Parts Symbol Part Name Code No. Specification Manufacturer WIB Operation display board LT00006042 Toho D
59 Replacement Parts Table BH600 Symbol Part Name Code No. Specification Manufacturer P Magnet pump 2-15-008-0013 MD30RZ-N Iwaki RF Compresso
60 Reference List of Dangerous Substances Never use explosive substances, flammable substances and substances that include explosive or flammable
61 Installation Standard Manual * Install the unit according the procedure described below (check options and special specifications separately).
62 Responsibility Please follow the instructions in this document when using this unit. Yamato Scientific has no responsibility for the accidents o
4 Before Using This Unit Requirements for Installation 1. Always ground this unit • Be sure to connect the earth wire (the green cable of power
5 Before Using This Unit Requirements for Installation 4. Do not modify • Modification of this unit is strictly prohibited. This could cause a fa
6 Before Using This Unit Requirements for Installation 8. Handling of power code • Do not entangle the power cord. This will cause overheating an